Bovie SurgiCenter PRO - 200 W Electrosurgical Generator A2350
Bovie SurgiCenter PRO - 200 W Electrosurgical Generator A2350
The Surgi-Center PRO is a 200 watt high power, operating room generator, by Bovie. The unit incorporates the power and safety of a premium OR generator at a significant savings to the medical facility. The system delivers 12 monopolar and bipolar RF energy modalities to provide a high level of surgical precision. Engineered with the safety of the patient in mind, the Bovie Surgi-Center|PRO includes technology such as BovieNEM™ return split pad contact quality monitoring, BovieFDFS™ tissue sensing technology, self-test circuits, as well as specialized modalities for laparoscopic and delicate procedures. The Bovie Surgi-Center|PRO is supported by Bovie’s unprecedented 4-year standard warranty.
Package Includes:
- Bovie Surgi-Center PRO
- Electrosurgical Generator
- 110 VAC Hospital-Grade Power Cord (15ft.)
- 220 VAC Hospital-Grade Power Cord (15ft.)
- User’s Guide in CD Format (MC-55-232-001)
Technical Specifications:
- Audible/Visible Output Meter: Yes
- Contact Monitoring Systems: BovieNEM™
- Contact Monitoring Systems Range/Trip Point: 10 ohms or less over entire operating range (10C° to 40°C)
- Cooling System: Passive Cooling
- Technology: Isolated RF Output
- Pre-set Feature: Yes
- Monopolar HP Ports: 2
- Dual Activation Tone Alert: Yes
- Connector Ports/Power Switch: White
- Weight: <9.07 kg (<20 lbs)
- Dimensions: W – 37.5 cm (14.75 in) H – 16.5 cm (6.5 in) D – 46 cm (18.1 in)
- Warranty: 4 years, Additional Warranty Available
MONOPOLAR | WATTS | VPEAK MAX VOLTAGE | CREST FACTOR (TYPICAL) | FREQUENCY (KHZ) |
---|---|---|---|---|
Cut I | 200 | 1000V | 1.7 ± 20% | 490 kHz ± 4.9 kHz |
Cut II | 200 | 750V | 1.7 ± 20% | 490 kHz ± 4.9 kHz |
Blend (1) | 200 | 1320V | 1.8 ± 20% | 490 kHz ± 4.9 kHz |
Blend (2) | 200 | 1475V | 2.0 ± 20% | 490 kHz ± 4.9 kHz |
Blend (3) | 200 | 1650V | 2.2 ± 20% | 490 kHz ± 4.9 kHz |
Blend (4) | 200 | 1870V | 2.4 ± 20% | 490 kHz ± 4.9 kHz |
Pinpoint Coagulation | 120 | 1800V | 3.1 ± 20% | 490 kHz ± 4.9 kHz |
Spray Coagulation | 120 | 4000V | 6.0 ± 20% | 350 to 450 kHz |
Gentle Coagulation | 120 | 450V | 1.6 ± 20% |
490 kHz ± 4.9 kHz |
BIPOLAR | WATTS | VPEAK MAX VOLTAGE | CREST FACTOR (TYPICAL) | FREQUENCY (KHZ) |
---|---|---|---|---|
Macro Bipolar | 80 | 600V | 1.5 ± 20% | 490 kHz ± 4.9 kHz |
Micro Bipolar | 80 | 500V | 1.5 ± 20% | 490 kHz ± 4.9 kHz |
Standard Bipolar | 50 | 250V | 1.5 ± 20% | 490 kHz ± 4.9 kHz |